Sensitive Terahertz Detectors: Room Temperature Terahertz Detectors based on Golay Cell.
|
Insight Product Company offers Room Temperature Terahertz (THz) Detectors
Room Temperature Terahertz Detectors: Golay Cells
Room Temperature Terahertz Detectors Type 2: Golay Cell
3. Golay Cell GC-1D
Golay Detector with Diamond Window
Golay detectors are manufactured and individually calibrated in-house and supplied from stock. Delivery includes a detector head and a power supply unit. A mount for the filters can be supplied as an option.
Due to polyethylene window exchange to Diamond one, GC-1D detectors have wider operation wavelength range spreading down to visible. They are usually used when someone needs not THz and VIS ranges only but MIR also. GC-1D model is a bit more expensive than GC-1T detector.
Various THz optical components such as low pass filters, polyethylene polarizers as well as windows, lenses, and beamsplitters from HRFZ-Si and TPX can be supplied as a useful complement for THz applications.
For transforming an analog signal of Golay cell to a digital signal we can offer hardware-software complex. It is served for detecting, processing, and analyzing optoacoustical detector signals. The complex consists of a specialized software and an electronic unit connecting Golay detector with personal computer through USB interface.
Application:
Monitoring and control VIS-THz radiation.
Technical specification:
Diameter of entrance cone, mm: |
11.0 |
Diameter of entrance window, mm: |
6.0 |
Material of entrance window: |
Diamond |
Operating wavelength range, μm: |
0.4 ÷ 8000 (Frequency range: 37.5 GHz - 750 THz) |
Recommended detected power, W, up to: |
1 x 10-5 |
Optimum modulation frequency, Hz: |
15 ± 5 |
Noise-equivalent power @ 20Hz:
typical, W/Hz1/2
minimum, W/Hz1/2 |
1.4 x 10^-10
0.8 x 10^-10
|
Optical responsivity @ 20Hz:
typical, V/W
maximum, V/W |
1 x 10^5
1.5 x 10^5
|
Response rate:
typical, ms
minimum, ms |
30
25
|
Detectivity (D*) at entrance cone aperture:
typical, cm x Hz1/2/W
maximum, cm |
7.0 x 10^9
11.0 x 10^9
|
Ambient operating pressure range, mm Hg |
760 ÷ 10-3 |
Operational and storage temperature range, degrees °C |
5 ÷ 40 |
Humidity, % |
no special requirements |
Vibration |
avoid vibrations at 1 - 100 Hz |
Rated voltage, VAC |
100/115 ± 10%,Â
220/230 ± 10% |
Line frequency, Hz |
50 - 60 |
Overall dimensions, L x W x H, mm3 |
126 x 45 x 87 |
Weight, kg |
0.8 |
|
Contact:
E-mail: |
admin@insight-product.com |
Phone: |
(617) 965-8151 |
|